(I)IoT Security News
ICS, News, Vulnerabilities

Baxter PrismaFlex and PrisMax (Update A)

Baxter PrismaFlex and PrisMax

1. EXECUTIVE SUMMARY

2. UPDATE INFORMATION

This updated advisory is a follow-up to the original advisory titled ICSA-20-170-01 Baxter PrismaFlex and PrisMax that was published June 18, 2020, on the ICS webpage on us-cert.gov.

3. RISK EVALUATION

Successful exploitation of these vulnerabilities may allow an attacker with network access to view and alter sensitive data. 

4. TECHNICAL DETAILS

4.1 AFFECTED PRODUCTS

The following models and versions of Baxter medical systems, are affected:

4.2 VULNERABILITY OVERVIEW

4.2.1    CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319

The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.

CVE-2020-12036 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:N/A:N).

4.2.2    IMPROPER AUTHENTICATION CWE-287

The affected devices do not require authentication when configured to send treatment data to a PDMS or an EMR system. This could allow an attacker to modify treatment status information.

CVE-2020-12035 has been assigned to this vulnerability. A CVSS v3 base score of 7.6 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:H/I:L/A:L).

4.2.3    USE OF HARD-CODED PASSWORD CWE-259

The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.

CVE-2020-12037 has been assigned to this vulnerability. A CVSS v3 base score of 5.4 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:L/UI:N/S:U/C:H/I:L/A:L).

4.3 BACKGROUND

4.4 RESEARCHER

Baxter reported these vulnerabilities to CISA.

5. MITIGATIONS

For PrismaFlex, Baxter recommends users update to Version SW 8.2x or later. The hard-coded service password vulnerability has been remediated in PrismaFlex Versions SW 8.2

For PrisMax, Baxter recommends users upgrade to PrisMaxv3 with DCM (Digital Communication Module), which supports mutually authenticated TLS tunnel to a PDMS or EMR system capable of implementing the latest TLS 1.2.

Additionally, Baxter recommends users of affected devices implement the following best practices:

Baxter also recommends that if a PDMS or EMR system is used with the affected devices, users should verify compatibility between the two systems. Users should also identify, analyze, evaluate, and control all risks associated with integration of medical devices in an enterprise network. Subsequent changes to the enterprise network could introduce new risks and require new analysis. The use of a PDMS or EMR system not compatible with the PrismaFlex and PrisMax systems can result in the presentation of erroneous data.

——— End Update A Part 1 of 1 ———

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:

CISA also provides a section for control systems security recommended practices on the ICS webpage on us-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.

No known public exploits specifically target these vulnerabilities.

Source:

https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02

Related posts

Siemens SIMATIC, SINAMICS, SINEC, SINEMA, SINUMERIK (Update B)

(I) IoT
4 years ago

Rockwell Automation Arena Simulation Software (Update B)

(I) IoT
5 years ago

Grundfos CIM 500

(I) IoT
4 years ago
Exit mobile version