(I)IoT Security News
ICS, News, Vulnerabilities

Capsule Technologies SmartLinx Neuron 2

Capsule Technologies SmartLinx Neuron 2

1. EXECUTIVE SUMMARY

2. RISK EVALUATION

Successful exploitation of this vulnerability could provide an attacker with full control of a trusted device on a hospital’s internal network.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of Capsule Technologies SmartLinx Neuron 2, a medical device platform, are affected:

3.2 VULNERABILITY OVERVIEW

3.2.1    PROTECTION MECHANISM FAILURE CWE-693

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running

Version 6.9.1. A specific series of keyboard inputs can allow an escape from the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

CVE-2019-5024 has been assigned to this vulnerability. A CVSS v3 base score of 7.6 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H).

3.3 BACKGROUND

3.4 RESEARCHER

Patrick DeSantis of Cisco Talos reported this vulnerability to Capsule Technologies.

4. MITIGATIONS

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on us-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.

This vulnerability is not exploitable remotely.

Source:

http://213.188.154.77/wp-admin/post.php?post=5398&action=edit