(I)IoT Security News
ICS, News, Vulnerabilities

Medtronic Conexus Radio Frequency Telemetry Protocol (Update C)

Medtronic Conexus Radio Frequency Telemetry Protocol

1. EXECUTIVE SUMMARY

2. UPDATE INFORMATION

This updated advisory is a follow-up to the advisory update titled ICSMA-19-080-01 Medtronic Conexus Radio Frequency Telemetry Protocol (Update B) that was published June 4, 2020, to the CISA webpage on us-cert.cisa.gov.

3. RISK EVALUATION

Successful exploitation of these vulnerabilities may allow an attacker with adjacent short-range access to one of the affected products to interfere with, generate, modify, or intercept the radio frequency (RF) communication of the Medtronic proprietary Conexus telemetry system, potentially impacting product functionality and/or allowing access to transmitted sensitive data. Successful exploitation requires: (1) an RF device capable of transmitting or receiving Conexus telemetry communication, such as a monitor, programmer, or software-defined radio (SDR); (2) to have adjacent short-range access to the affected products; and (3) for the products to be in states where the RF functionality is active. Before the device implant procedure and during follow-up clinic visits, the Conexus telemetry sessions require initiation by an inductive protocol. Outside of these use environments, the RF radio in the affected implanted device is enabled for brief periods of time to support scheduled follow-up transmissions and other operational and safety notifications. The result of successful exploitation of these vulnerabilities may include the ability to read and write any valid memory location on the affected implanted device and therefore impact the intended function of the device.

4. TECHNICAL DETAILS

4.1 AFFECTED PRODUCTS

The following products and versions of Medtronic devices utilizing the Conexus telemetry protocol are affected:

Not all devices may be available in in all geographical locations.

4.2 VULNERABILITY OVERVIEW

4.2.1    IMPROPER ACCESS CONTROL CWE-284

The Conexus telemetry protocol utilized within this ecosystem does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.

CVE-2019-6538 has been assigned to this vulnerability. A CVSS v3 base score of 9.3 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H).

4.2.2    CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319

The Conexus telemetry protocol utilized within this ecosystem does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

CVE-2019-6540 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N).

4.3 BACKGROUND

4.4 RESEARCHER

Peter Morgan of Clever Security; Dave Singelée and Bart Preneel of KU Leuven; Eduard Marin formerly of KU Leuven, currently with University of Birmingham; Flavio D. Garcia; Tom Chothia of the University of Birmingham; and Rik Willems of University Hospital Gasthuisberg Leuven reported these vulnerabilities to CISA.

5. MITIGATIONS

Medtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval.

Medtronic has developed mitigating patches for a subset of the affected implanted cardiac device models. These patches are installed during regular office visits. Medtronic has stated that patches for additional impacted models are being developed by Medtronic and will be deployed through future updates. Patches are currently available for the following affected models:

——— Begin Update C Part 1 of 1 ——– 

——— End Update C Part 1 of 1 ——– 

Medtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:

Medtronic has released additional patient focused information.

CISA recommends users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on us-cert.cisa.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.cisa.gov in the Technical Information Paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.

These vulnerabilities are not exploitable remotely.

Source:

https://us-cert.cisa.gov/ics/advisories/ICSMA-19-080-01

Related posts

VMware Releases Security Advisory for Aria Operations

IoT
10 months ago

Open Design Alliance Drawings SDK

(I) IoT
4 years ago

Reliable Controls MACH-ProWebCom/Sys

(I) IoT
5 years ago
Exit mobile version