(I)IoT Security News
ICS, News, Vulnerabilities

Philips Patient Monitoring Devices

Philips Patient Monitoring Devices

1. EXECUTIVE SUMMARY

2. RISK EVALUATION

Successful exploitation of these vulnerabilities could result in unauthorized access, interrupted monitoring, and collection of access information and/or patient data.

Note: To successfully exploit these vulnerabilities, an attacker would need to gain either physical access to surveillance stations and patient monitors or access to the medical device network.

3. TECHNICAL DETAILS

3.1 AFFECTED PRODUCTS

The following versions of the patient monitoring devices are affected:

3.2 VULNERABILITY OVERVIEW

3.2.1    IMPROPER NEUTRALIZATION OF FORMULA ELEMENTS IN A CSV FILE CWE-1236

The software saves user-provided information into a comma-separated value (CSV) file, but it does not neutralize or incorrectly neutralizes special elements that could be interpreted as a command when the file is opened by spreadsheet software. 

CVE-2020-16214 has been assigned to this vulnerability. A CVSS v3 base score of 4.2 has been calculated; the CVSS vector string is (AV:L/AC:L/PR:H/UI:R/S:C/C:L/I:L/A:N).

Affects the following products: Patient Information Center iX (PICiX) Versions B.02, C.02, C.03.

3.2.2    IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION (‘CROSS-SITE SCRIPTING’) CWE-79

The software does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output that is then used as a webpage and served to other users. Successful exploitation could lead to unauthorized access to patient data via a read-only web application. 

CVE-2020-16218 has been assigned to this vulnerability. A CVSS v3 base score of 3.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:L/UI:N/S:U/C:L/I:N/A:N).

Affects the following products: Patient Information Center iX (PICiX) Versions B.02, C.02, C.03.

3.2.3    IMPROPER AUTHENTICATION CWE-287

When an actor claims to have a given identity, the software does not prove or insufficiently proves the claim is correct. 

CVE-2020-16222 has been assigned to this vulnerability. A CVSS v3 base score of 5.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L).

Affects the following products: Patient Information Center iX (PICiX) Version B.02, C.02, C.03, and PerformanceBridge Focal Point Version A.01.

3.2.4    IMPROPER CHECK FOR CERTIFICATE REVOCATION CWE-299

The software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate. 

CVE-2020-16228 has been assigned to this vulnerability. A CVSS v3 base score of 6.0 has been calculated; the CVSS vector string is (AV:A/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:L).

Affects the following products: Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior.

3.2.5    IMPROPER HANDLING OF LENGTH PARAMETER INCONSISTENCY CWE-130

The software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart. 

CVE-2020-16224 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

Affects the following products: Patient Information Center iX (PICiX) Versions C.02, C.03.

3.2.6    IMPROPER VALIDATION OF SYNTACTIC CORRECTNESS OF INPUT CWE-1286

The product receives input that is expected to be well-formed (i.e., to comply with a certain syntax) but it does not validate or incorrectly validates that the input complies with the syntax, causing the certificate enrollment service to crash. It does not impact monitoring but prevents new devices from enrolling. 

CVE-2020-16220 has been assigned to this vulnerability. A CVSS v3 base score of 3.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:R/S:U/C:N/I:N/A:L).

Affects the following products: Patient Information Center iX (PICiX) Versions C.02, C.03, PerformanceBridge Focal Point Version A.01.

3.2.7    IMPROPER INPUT VALIDATION CWE-20

The product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart. 

CVE-2020-16216 has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been calculated; the CVSS vector string is (AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).

Affects the following products: IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior.

3.2.8    EXPOSURE OF RESOURCE TO WRONG SPHERE CWE-668

The product exposes a resource to the wrong control sphere, providing unintended actors with inappropriate access to the resource. The application on the surveillance station operates in kiosk mode, which is vulnerable to local breakouts that could allow an attacker with physical access to escape the restricted environment with limited privileges. 

CVE-2020-16212 has been assigned to this vulnerability. A CVSS v3 base score of 6.8 has been calculated; the CVSS vector string is (AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).

Affects the following products: Patient Information Center iX (PICiX) Versions B.02, C.02, C.03.

3.3 BACKGROUND

3.4 RESEARCHER

Julian Suleder, Nils Emmerich, Birk Kauer of ERNW Research GmbH, Dr. Oliver Matula of ERNW Enno, and Rey Netzwerke GmbH reported these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices), which reported these to Philips.

4. MITIGATIONS

Philips plans a new release to remediate all reported vulnerabilities:

As a mitigation to these vulnerabilities, Philips recommends the following:

Users with questions regarding their specific Philips Patient Information Center (PIC iX) and/or IntelliVue patient monitor installations and new release eligibility should contact their local Philips service support team, or regional service support, or call 1-800-722-9377.

Please see the Philips product security website for the Philips advisory and the latest security information for Philips products.

CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:

CISA reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.

CISA also provides a section for control systems security recommended practices on the ICS webpage on us-cert.gov. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.

Additional mitigation guidance and recommended practices are publicly available on the ICS webpage on us-cert.gov in the Technical Information Paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies.

Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to CISA for tracking and correlation against other incidents.

CISA also recommends users take the following measures to protect themselves from social engineering attacks: 

No known public exploits specifically target these vulnerabilities. These vulnerabilities are not exploitable remotely.


Source:

https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01